FTIR/FT-NIR module Interspec 400-X


Specifications
Wavelength range, IR 7000 to 400 cm-1
Wavelength range, NIR 15000 to 3000 cm-1
Resolution, standard 2 cm-1 (8 cm-1 - NIR)
Resolution, option 1.0 or 0.5 cm-1
Interferometer Self aligning
Output beam diameter 38 mm
Aperture ratio f 3.2
Beamsplitter, standard IR Multicoated KBr
Beamsplitter, option ZnSe, CaF2, CsI
Beamsplitter, standard NIR Fused Silica
Frequency reference HeNe laser
IR source High intensity air cooled
NIR source Quartz-halogen lamp
Dessiccant possibility Yes
Interface Arcnet
Power 2.35 VDC (5.5A), +/-12 V, +5 V, 35 W
Dimensions W33xD25xH16 cm
Weight 11 kg
Temp. environment 15 - 28 0C
Humidity environment Best below 65%


General

The INTERSPEC series FTIR modules are stable, rugged units requiring the minimum of external utilities.

*The sealed, desiccated enclosure keeps the interferometer dry

*The IR/NIR source does not require water cooling

*The interferometer mechanical bearing does not require a gas supply.

For easy access, those system components which have a finite lifetime (i.e. laser, and radiation source) are located at the front side of the module. A yellow LED indicator indicates, when the scanning is turned on.

Interferometer Performance

The interferometer compartment under cover is 250 x 250 mm wide and 140 mm high. The Interspec 400-X series of instruments employ a sealed and desiccated interferometer, ensuring high spectral integrity with low levels of water vapour within the interferometer. We also offer a version (400-XZ) that employs a ZnSe moisture insensitive optics to be impervious to water vapour and can be used to advantage in serious tropical environment's.

Near infrared version (400-XN) employs a fused silica optics and is insensitive to any influence of water vapour.

Resolution in the infrared is available factory default from 2 cm-1 to 32 cm-1 (option 0.5 and 1 cm-1). The overall wavelength range is 7000 to 400cm-1 (IR) or 15000 to 3000 cm-1 (NIR).

Extending Wavelength Ranges

In order to facilitate the use of more than one beam splitter or detector , provision has been made to interchange the beam splitter and detector assemblies allowing the Interspec 400-X to be used at any wavelength from 15000 to 400 cm-1.


Beam Splitters Range subject to coatings
KBr 7,000 to 400 cm-1
ZnSe 5,000 to 500 cm-1
CaF2 10,000 to 1,000 cm-1
Quartz 15,000 to 3,000 cm-1

Sample Handling Accessories

Interspec 400-X series instruments may be used as an heart of a more complex spectroscopic system. Sampling handling and the whole system is defined by the customer.

Models

Interspec 400-X - Standard FTIR Portable Instrument with KBr optics.
Interspec 400-XZ - Non Water Sensitive FTIR Portable Instruments for High Humidity with ZnSe optics
Interspec 400-XN - Near Infra Red Portable Instrument with fused silica optics.